M19 Micropore Analyzer for technical textiles — bubble point, mean flow pore, and pore size distribution to ASTM D6767 & ASTM F316. Capillary flow porom...
Frequently asked questions
What is the difference between bubble point and mean flow pore size in technical textiles?
The bubble point represents the largest through-pore in the fabric, measured at the pressure where the first continuous gas bubble breaks through the wetted sample. It determines the absolute filtration rating and barrier protection limit. The mean flow pore (MFP) is the pore diameter at which 50% of the continuous gas flow passes through smaller pores and 50% through larger pores, representing the nominal filtration rating.
How does ASTM D6767 porometry compare to traditional dry glass bead sieving (ASTM D4751) for geotextiles?
Traditional dry sieving with glass beads is labor-intensive, takes hours, suffers from electrostatic bead entrapment, and alters fabric weave during shaking. ASTM D6767 capillary flow porometry provides an operator-independent, non-destructive test in under 5 minutes, delivering the complete continuous pore size distribution curve and accurate O95 opening size.
Can the MPA test both open woven fabrics and lofty needle-punched felts without crushing them?
Yes. The M19 MPA uses specialized pneumatic sample chambers with calibrated torque clamping and optional support screens. This prevents edge gas bypass leakage while ensuring delicate or lofty nonwovens are not compressed, preserving their native pore structure during measurement.
Why is Capillary Flow Porometry preferred over Mercury Intrusion Porosimetry (MIP) for textiles?
Mercury porosimetry requires pressures up to 60,000 psi, which physically crushes and destroys flexible textile fibers. It also measures non-functional dead-end pores and uses hazardous, toxic mercury. Capillary flow porometry uses safe, non-toxic wetting fluids at low pressures and exclusively characterizes active through-pores that carry actual air or liquid flow.
Is capillary flow porometry the same as a Frazier air permeability test?
No. Air permeability gives only a single bulk flow number at a fixed pressure drop, but cannot tell you maximum pore size, pore uniformity, or pinhole defects. Porometry measures the full distribution of pore diameters that explains the permeability and predicts particulate capture and barrier performance.